Precautions and Responsibilities

Precautions specified in IEC 60825-1 and FDA CFR Title 21 Part 1040 are as follows:

Requirement

Class 2

Class 2M

Class 3R

Class 3B

Remote interlock

Not required

Not required

Not required

Required*

Key control

Not required

Not required

Not required

Required – cannot remove key when in use*

Power-on delays

Not required

Not required

Not required

Required*

Beam attenuator

Not required

Not required

Not required

Required*

Emission indicator

Not required

Not required

Not required

Required*

Warning signs

Not required

Not required

Not required

Required*

Beam path

Not required

Not required

Terminate beam at useful length

Terminate beam at useful length

Specular reflection

Not required

Not required

Prevent unintentional reflections

Prevent unintentional reflections

Eye protection

Not required

Not required

Not required

Required under special conditions

Laser safety officer

Not required

Not required

Not required

Required

Training

Not required

Not required

Required for operator and maintenance personnel

Required for operator and maintenance personnel

*LMI Class 3B laser components do not incorporate these laser safety items. These items must be added and completed by customers in their system design. For more information, see Class 3B Responsibilities.